When an alternating pressure mattress loses power, it stops active cycling; however, the surface may still be functional for some time. How long this can continue depends on several factors: the construction of the mattress, the use of non-return valves, how the hoses are connected to the pump, whether there is a transport cap/clamp available, if there is a foam base beneath the air cells, whether there is a battery back-up system, what weight is being placed on the mattress, and how long the power interruption will last. A safe course of action would include having a specific plan for your particular model that addresses observation of the patient, handling of the hoses during transportation, resetting the mattress, etc.
While power loss and patient transportation have some similarities, they are distinct. When a power loss occurs, the focus is on providing a suitable support surface and observing for changes. In addition to preparing for the move (the mattress, pump, cables, and hoses), static mode, transport mode, battery operation, and CPR deflation each serve different purposes and cannot be used interchangeably.
Conoce el sistema antes de que se produzca un corte de suministro eléctrico
Un plan de corte de suministro eficaz debe partir de las instrucciones exactas del colchón y la bomba. Las familias de productos pueden comportarse de forma diferente, aunque sus paneles de control parezcan similares.
| Detalles del sistema | Qué hay que confirmar | Por qué cambia la respuesta |
|---|---|---|
| Retención de aire | Cómo varía la presión una vez que la bomba se detiene | Determina la rapidez con la que puede reducirse la asistencia |
| Conector para manguera | Independientemente de si contiene válvulas antirretorno | Algunos conectores retienen mejor el aire cuando se dejan colocados |
| Dispositivo de transporte | Tapa, tapón, abrazadera, posición del dial o ninguno | Define el método autorizado para limitar la pérdida de aire durante el transporte |
| Base de espuma | Presencia, solidez y función prevista como respaldo | Podría ofrecer un soporte subyacente a medida que descienda la presión atmosférica |
| Batería o alimentación de reserva | Tipo homologado, tiempo de funcionamiento, indicación de carga y carga | Puede permitir que se siga bombeando durante un periodo determinado |
| Alarma y memoria | Alerta de corte de corriente y comportamiento ante el reinicio | Indica si es necesario volver a introducir los ajustes |
| Límite de transporte | Duración permitida y condiciones | Evita que una situación de transporte temporal se convierta en una práctica habitual |
Mantenga a mano, junto al expediente del equipo, las instrucciones del modelo, la identificación de la bomba, el modo normal de funcionamiento, los ajustes habituales, los contactos en caso de avería y el plan de asistencia alternativo. Una breve prueba controlada permite comprobar el comportamiento de las alarmas y el reinicio cuando esté autorizado y el colchón no tenga a ningún usuario sobre él.
Prepárate para un corte de suministro eléctrico programado
Planned electrical work, room changes, bed maintenance, or scheduled transportation will give you time to plan. You should find out how much time the electricity will be off and then see how long the manufacturer says the item can travel or retain air for during that time.
Lista de comprobación para la preparación
- Identifica la combinación exacta de bomba y colchón y lee las instrucciones relativas a la pérdida de potencia.
- Comprueba el modo de funcionamiento actual y el ajuste de presión o de confort.
- Comprueba si la bomba incluye una batería homologada y cargada o una fuente de alimentación de reserva externa homologada.
- Localiza la posición específica del tapón de transporte, el tapón, la abrazadera o el conector del modelo, si se incluye alguno.
- Comprueba que la válvula de control de la RCP esté cerrada y que siga siendo accesible.
- Revisa las mangueras, las celdas, la cubierta, el cable de alimentación y los conectores antes de la interrupción.
- Prevé una superficie de apoyo alternativa o un plan de traslado si la interrupción pudiera prolongarse más allá de lo autorizado.
- Decide quién se encargará de vigilar la superficie, restablecer el suministro eléctrico y comprobar que el sistema se reinicia.
No des por sentado que un sistema de alimentación ininterrumpida (SAI) de uso general sea adecuado. Son factores importantes la potencia nominal, la forma de onda de salida, la conexión a tierra, el comportamiento de las alarmas, el tiempo de autonomía, la inspección y la normativa local sobre equipos. Utiliza únicamente una fuente de alimentación de reserva homologada para el dispositivo y el entorno en cuestión.
Cómo actuar ante un corte de luz inesperado
First determine if the electrical problem is affecting only the outlet, the entire room, or the rest of the building. Make sure the pump’s power cord is plugged in and check for damage on the cord. If there are any other indications of a loss of electricity such as visual or audible indicators, do not switch the control repeatedly to verify them.
- Comprueba quién es el usuario, la forma del colchón, la posición de la cama y el estado del soporte inmediato.
- Anota la hora en la que se detuvo el bombeo activo o en la que se comprobó que se había detenido.
- Mantén cerrada la válvula de liberación de CPR y evita desconectar las mangueras, a menos que así lo exija el procedimiento de corte de suministro del modelo.
- Sigue los pasos de transporte o retención de aire aprobados para ese diseño concreto de conector.
- Reduzca los movimientos innecesarios, los cambios bruscos de posición en la cama y cualquier cosa que comprima la manguera o el colchón.
- Ponga en marcha el plan alternativo de suministro o de transferencia si el suministro está disminuyendo o si no se va a restablecer dentro de los plazos permitidos.
- Continuar con la observación hasta que se restablezca y se compruebe el funcionamiento normal, o hasta que el usuario reciba asistencia en otro lugar.
Consejo práctico: Anota la hora de inicio de la pérdida de firmeza en un papel o en el historial de cuidados tan pronto como se conozca. Esto es más fiable que intentar estimar la duración una vez que el colchón ya se ha ablandado.
Observa la superficie del colchón mientras la bomba está apagada
A non-audible pump is no longer adjusting pressure or performing an alternating cycle. The mattress can continue to hold its air, gradually release its air, or lose air very fast through a leak in the connector or cell. As such, the focus of observations should be on the overall state of the support surface, as opposed to just whether the pump is making noise.
Cambios a tener en cuenta
- Una reducción apreciable de la altura del colchón o una alteración de su perfil habitual.
- El usuario se hunde más o se acerca a la base de la cama.
- Un grupo de celdas pierde aire más rápido que el resto del colchón.
- Una manguera, un conector, un mando de RCP o una cubierta que cambian de posición a medida que se mueve la cama.
- Menor estabilidad de los bordes durante el reposicionamiento o la preparación para el traslado.
- Un cambio en la comodidad, la postura, la alineación o la capacidad para moverse con seguridad.
When possible use the model’s specified base support or bottom-out check. Do not push through the cover forcefully, nor rely on a single hand check if you are required to use another method. The frequency of repositioning, timing of transfers, and use of alternate surfaces must be based on the individual’s current care plan and observed status.
Maneja las mangueras de aire de acuerdo con el diseño de transporte
El manejo de la manguera determina la cantidad de aire que se escapa de muchos sistemas. En algunos colchones, dejar el conector acoplado a una bomba apagada ayuda a retener el aire. Otros sistemas indican al usuario que desconecte el conector y coloque inmediatamente un tapón de transporte, una los extremos de la manguera, cierre un dial de transporte o aplique una abrazadera específica.
- Utiliza únicamente el accesorio de transporte suministrado o especificado para ese colchón.
- No improvises haciendo nudos en el tubo, doblándolo con fuerza ni sujetándolo con herramientas comunes.
- No abras nunca el control de RCP para preparar un traslado rutinario.
- Mantén limpios y protegidos los extremos de las mangueras que estén tapados o desconectados.
- Evita colocar el conector debajo del colchón o en un lugar donde pueda recibir un golpe del armazón de la cama.
- Tras el transporte, restablezca la configuración normal de las mangueras antes de volver a poner en marcha la bomba.
Una funda de transporte limita la pérdida de aire únicamente cuando las celdas circundantes, las mangueras, las juntas y el cierre CPR están intactos. No convierte cualquier colchón en una superficie no motorizada de larga duración.
Preparación del colchón para el transporte de pacientes en trayectos cortos
A transport cap will prevent air from leaking out of the mattress, but only if all of the components that make up the mattress (i.e., cells, hoses, seals, CPR closure) are functioning correctly. A transport cap is not able to convert each mattress into a non-powered surface with extended durability.
Cuando el usuario se desplaza sobre la cama
- Si el tiempo lo permite, deja que el colchón alcance su estado normal de funcionamiento antes de desconectarlo de la corriente.
- Aplica la configuración de transporte específica del modelo, la tapa, el conector o el procedimiento homologado para la batería.
- Fija la bomba en su posición homologada o colócala en el lugar previsto para su transporte.
- Enrolla y fija el cable de alimentación de manera que no pueda entrar en contacto con las ruedas, el suelo, los rieles ni los marcos de las puertas.
- Coloca las mangueras dejando suficiente holgura para que la cama pueda moverse, pero sin que queden bucles sueltos ni puntos en los que puedan pellizcarse.
- Comprueba que el mando de la RCP siga estando accesible y cerrado.
- Observa la altura del colchón y la posición del usuario durante las pausas y nada más llegar.
Do not use the power cable as a hanging point for the pump, and avoid allowing it to move freely while being carried via the hose connection. In cases where the pump will not stay firmly in place during travel, you must only utilize the approved storage/travel methods outlined within the system’s documentation.
Cuando se traslada al usuario a una camilla o a una silla
The transfer will depend upon what type of handling equipment has been selected and the assistance plan. You need to determine if the air mattress stays in place on the bed; travels with the user as a viable transfer surface; or if a different support surface replaces the air mattress at the final destination. A standard alternating mattress can not function as an air-assisted transfer mattress unless it was specifically designed to do so.
When the user gets out of bed, ensure that the identification of the mattress and pump setup is maintained. Additionally, prevent any loose cables/hoses from becoming a safety hazard (trip) and do not leave the system inflated fully, in transport mode, or disconnected from power supply unless there is a clear label indicating its status.

Reiniciar el sistema como comprobación de puesta en servicio
When you are ready to turn on the power, or when the bed has reached where it is supposed to be, then you need to plug it back in. This will be your initial start-up process. As such, consider restarting as a quick check-out to make sure that nothing was disturbed during travel; i.e., whether the settings on the unit, the positioning of the hoses, the closing of the CPR valve, and the pressure of the mattress has changed.
- Coloca la plataforma en la posición indicada para el reinicio y acciona los frenos.
- Retira la tapa o abrazadera de transporte y vuelve a conectar cada manguera a su conexión habitual.
- Comprueba que los conectores estén bloqueados, que las mangueras estén abiertas y que la válvula de control de la RCP esté cerrada.
- Revisa el cable de alimentación y conecta la bomba a la toma de corriente homologada.
- Observa los indicadores de autocomprobación al arrancar, de pantalla y de alarma.
- Vuelve a introducir o comprueba el modo y el ajuste de presión deseados; no des por hecho que la memoria los haya guardado.
- Deja que el colchón alcance su estado óptimo y observa la secuencia celular prevista.
- Comprueba el soporte, la posición del usuario, el movimiento de la camilla, el recorrido de las mangueras y el estado de las alarmas antes de reanudar el uso habitual.
While a soft mattress is inflating again, the low pressure alarm can stay on. This will depend upon the specific model’s startup sequence. When the alarm continues to ring after the soft mattress has been re-inflated, check the following areas prior to putting the system back into service: Transport Device, Hose Connection, CPR Valve, Cells, Filter and Pump Output.
Registrar el suceso y restablecer la disponibilidad
A quick record will assist you in explaining future changes to your mattress pressure, settings, alarms, or equipment status. You should make note of the time when the interruption began and ended, which mattress and pump were being used, if there was transportation involved, what the hoses or caps look like, if back up power was utilized, any observations made about the surface, the reaction of the user, the restart settings, the alarms, and the final test results.
Al revisar sistemas de colchones de presión alterna, busca instrucciones claras sobre cortes de corriente, componentes de transporte, comportamiento de las alarmas, memoria de ajustes, información sobre baterías compatibles (si está disponible) y una lista de comprobación para el reinicio. Estos detalles influyen en la facilidad de uso diaria tanto como el ciclo normal de alternancia.
Comprobaciones del equipamiento tras el evento
- Recarga una batería homologada y anota cualquier disminución de la autonomía o cualquier aviso.
- Devuelve las tapas, las abrazaderas y los accesorios de transporte a su lugar correspondiente.
- Comprueba que los conectores y los cables no presenten daños causados durante el traslado.
- Limpia cualquier componente que haya estado en contacto con el suelo, el ascensor, la superficie del pasillo o el equipo de transporte.
- Notifica cualquier alarma persistente, pérdida rápida de aire, fallo en el reinicio o cambio en el sonido de la bomba.
- Actualiza el plan de corte de suministro si el incidente real ha puesto de manifiesto que faltan instrucciones o equipos.
Preguntas frecuentes
¿Cuánto tiempo permanecerá inflado un colchón de presión alterna sin corriente eléctrica?
The amount of time the air is held in the mattress will vary depending upon the type of mattress being used, the type of valve used, where the connector is located, what type of transport device is used, whether there are leaks in the cells, how much weight is applied to the mattress, and what position the bed is placed in. Use the specified times and conditions that apply to your particular model.
¿Hay que desconectar las mangueras de aire durante un corte de corriente?
Solo si las instrucciones del producto indican que hay que desconectarlos y explican cómo mantener el aire después. Algunos sistemas retienen mejor el aire mientras están conectados; otros utilizan una tapa de transporte, un tapón, un dial o una abrazadera.
¿Sirve el modo estático como sistema de respaldo en caso de corte de corriente?
No. Static mode will still require a running pump for the maintenance of a controlled pressure. When the system loses power, there will be no active pressure control or cycling.
¿Se puede utilizar la válvula de RCP como control de transporte?
The purpose of the CPR control is to quickly deflate the mattress. During routine transport, the method used with the mattress (i.e. leaving the connector plugged in or using the designated transport accessory) should be followed.
Qué hay que comprobar una vez que se restablezca el suministro eléctrico
Comprueba las mangueras, los componentes de transporte, el cierre del sistema de reanimación cardiopulmonar (RCP), el cable de alimentación, los indicadores de puesta en marcha, el modo, el ajuste de presión, el inflado, la secuencia de las celdas, la asistencia al usuario, el movimiento de la camilla y las alarmas no resueltas.
Conclusión
To manage an alternating pressure mattress when you experience power loss, you need to be prepared before the power goes out. This preparation includes understanding how the specific mattress retains air, what kind of connector or transport device it uses, if there is a backup power source that has been approved for use with your particular model, and when you should initiate an alternative transfer strategy.
During an outage, you should focus on the surface and user instead of only focusing on the silent pump. When transporting, be sure to secure the pump, cable, and hoses. Do not use the CPR release as a transport control. After power has returned, restore the normal air path; verify settings; allow reinflation; check the operating cycle; and document the final status.